Validation of a single-platform, volumetric, flow cytometry for CD4 T cell count monitoring in therapeutic mobile unit

Abstract

Background

A mobile health unit may be useful to follow up adult and pediatric patients on antiretroviral treatment and living in remote areas devoid of laboratory facilities. The study evaluated the use of the simplified, robust, single-plateform, volumetric, pan-leucogating Auto40 flow cytometer (Apogee Flow Systems Ltd, Hemel Hempstead, UK) for CD4 T cell numeration in a mobile unit, compared against a reference flow cytometry method.

Methods

The therapeutic mobile unit of the Laboratoire National de Santé Hygiène Mobile, Yaoundé, Cameroon, was equipped with the Auto40. A FACSCalibur flow cytometer (Becton Dickinson Immuno-cytometry System, San Jose, CA, USA) was used as reference method. EDTA-blood samples from volunteers were first subjected to CD4 T cell count in the mobile unit, and an aliquot was sent within 4 hours to Centre International de Référence Chantal Biya, Yaoundé, for FACSCalibur assay.

Results

Two HIV screening campaigns with the mobile unit were organised in December 2009 and January 2010. The campaign in the suburb of Yaoundé which was 20 km from the reference laboratory included 188 volunteers comprising 93 children less than 5 years old. The campaign in Ambang Bikok (53 km far from Yaoundé) included 69 adult volunteers. In Yaoundé suburb, mean ± standard deviation (SD) CD4 T cell count was 996 ± 874 cells/μl by Auto40, and 989 ± 883 cells/μl by FACSCalibur; in Ambang Bikok, mean ± SD CD4 T cell count was 1041 ± 317 cells/μl by Auto40, and 1032 ± 294 cells/μl by FACSCalibur. Results by Auto40 and FACSCalibur were highly correlated in Yaoundé (r2 = 0.982) as in Ambang Bikok (r2 = 0.921). Bland-Altman analysis showed a close agreement between Auto40 and FACSCalibur results expressed in absolute count as in percentage in Yaoundé and Ambang Bikok. When pooling the 257 CD4 T cell count measurements, the Auto40 yielded a mean difference of +7.6 CD4 T cells/μl higher than by reference flow cytometry; and the sensitivity and specificity of Auto40 in enumerating absolute CD4 T cell counts of less than 200 cells/μl were 87% and 99%, respectively, and in enumerating absolute CD4 T cell counts of less than 350 cells/μl were 87% and 98%, respectively. The intrarun and interun precisions of the Auto40 assay assessed in the mobile unit were 5.5% and 7.9%, respectively.

Conclusions

The Auto40 flow cytometer installed in a therapeutic mobile unit and operated far from its reference laboratory gave a perfect correlation with the reference method, and could be useful in carrying out immunological monitoring of HIV-infected patients living in areas without access to laboratory facilities.

Background

The 2010-revised guidelines of the World Health Organization (WHO) for scaling up of antiretroviral treatment (ART) in adults and children living in resource-limited settings [1, 2], emphasizes the need of laboratory monitoring, based first on immunological assessment by the numeration of CD4 T lymphocytes, mainly to start ART and monitor patients on ART, and secondly on HIV-1 RNA load in order to monitor treatment efficacy, early therapeutic failure and subsequent therapeutic switch [36]. Affordable CD4 T cell counting has gradually become possible by using simple, compact and robust low-cost new generation cytometers operating as single-platform volumetric instruments without the use of expensive micro beads [711].

The scaling up of public ART programs globally has led to an increased demand for CD4 T cell count tests [12], especially to assess treatment eligibility. CD4 T cell count has become the most decentralized and universally available biological marker to monitor HIV disease and ART [3]. However, access to CD4 T cell measurement remains a bottleneck to ART scale-up and decentralization, especially in remote rural areas frequently lacking laboratory infrastructure [13, 14].

In order to increase the number of individuals tested for HIV, in keeping with our target of universal access, we recently developed a strategy based on bringing the healthcare package much closer to the people, through mobile HIV testing units [15], conceived to bring education and testing services to hard-to-reach population [16]. This pilot study, the first in the Central African region, clearly demonstrated the acceptability, feasibility and effectiveness of using mobile units as a tool for mass HIV testing in individuals with limited access to voluntary counseling and testing [15]. Successful HIV testing with mobile units was unexpected, given that resistance and sociocultural impediments to HIV testing have been identified in Central Africa [17]. In addition, previous successful experiences in South Africa [18] and Kenya [19] showed that mobile services for HIV screening are accessed by different target population compared with facility-based services, including younger people, and higher proportion of persons with newly diagnosed HIV infection.

HIV counseling and screening mobile units may be complemented with antiretroviral drugs and laboratory facilities to become ambulatory health units for ART and follow up adult and pediatric patients living in remote areas devoid of laboratory facilities. The recently developed Auto40 flow cytometer (Apogee Flow Systems Ltd, Hemel Hempstead, UK; http://www.ApogeeFlow.com) was originally developed for military applications [10]. It has been adapted for CD4 T cells measurement within 30 minutes, using a pan-leucogating protocol with anti-CD4 and anti-CD45 thermo-stable monoclonal antibodies. Due to the stability of its optical bench, the Auto40 has been conceived for use on mobile or peripheral stationary flow cytometry unit [10]. Finally, the aim of the present study was to evaluate the usefulness of the simplified Auto40 flow cytometry system for CD4 T cell numeration in mobile health unit, according to a reference flow cytometry method.

Materials and methods

Volunteers, blood sampling and processing

In December 2009 and in January 2010, two HIV screening campaigns with mobile units were organized in the outskirts of Yaoundé, the capital city of Cameroon, 20 km away from the Laboratoire National de Santé Hygiène Mobile (LNSHM), the reference laboratory for HIV screening mobile campaign in Cameroon [15], and in Ambang Bikok, a rural setting located 53 km from Yaoundé. In a suburb of Yaoundé, children followed up at the Nkoldongo Medical Health District were also included for CD4 T cell measurement, as part of an agreement with the Chantal Biya International Reference Centre for Research and Prevention of HIV/AIDS (CIRCB), Yaoundé, for free HIV biological monitoring. Volunteers had access to primary health care centers and antenatal clinics, but modern laboratory facilities were not routinely available.

An aliquot of K3-EDTA-blood sample obtained by venipuncture in Vacutainer tubes (Becton Dickinson, Franklin Lakes, NJ, USA), kept at ambient temperature, was obtained after informed consent from each volunteer, patient or child guardian. Each aliquot was first subjected to CD4 T cell count in the therapeutic mobile unit, and a second aliquot was send within 4 hours at CIRCB for measurement by flow cytometry reference analyzer in the same day of sampling. All blood samples were unlinked to identifiers.

CD4 T cell measurements

CD4 T cell counting was performed on 2 different flow cytometers: (1) the FACSCalibur [Becton Dickinson Immuno-cytometry System (BDIS), San Jose, CA, USA], a dedicated clinical instrument for CD4 T cell counting installed at the CIRCB, used as reference method, and (2) the Auto40 flow cytometer (Apogee Flow Systems Ltd) equipped with a green laser at 532 nm, a side scatter detector, two fluorescence channels and means for direct volumetric counting, without requiring a step of red blood cell lysis. The Auto40 flow cytometer has been installed in the therapeutic mobile unit of the LNSHM, using the battery of the van as the source of electrical energy. An optional sheath fluid recycling cassette reduces sheath fluid consumption and may be of interest in remote areas where the supply is limited.

The FACSCalibur was used as predicate instrument for CD4 T cell counting in the study, as a single-platform flow cytometry technique. Absolute CD4 T cell counting was performed according to the instructions provided by the manufacturer. In brief, 50 μl of EDTA-blood and 10 μl of Multitest reagent (BDIS) were added to the dedicated TruCOUNT™ tubes (BDIS) containing a fixed number of polystyrene reference beads and then vortexed and left to incubate for 15 minutes at room temperature. The Multitest reagent is combined with four fluorescent [fluorescein isothiocyanate (FITC), phycoerythrin (PE), peridinin chlorophyll protein (PerCP), and alophycocianin (APC)]-conjugated monoclonal antibodies reagents for CD3/CD8/CD45/CD4 (FITC-CD3/PE-CD8/PerCP-CD45/APC-CD4). Red blood cells were then lysed by using FACS Lysing Solution (BDIS). The lyse-no-wash stained samples were run on the FACSCalibur. Data were analyzed using MultiSet V2.2 software (BDIS) for calculating the absolute and percentage values of CD4 T cells. The CD8 T cell counting allowed by the Multitest reagent was not used in the present study

The Auto40 assay is based on the no lyse procedure [20], which avoids the red blood cell lysis step, thereby reducing assay variability due to changes in assay conditions (time and temperature of incubation) as well as differences in the susceptibility of cells to the lysis reagents [21]. The Auto40 analyzer uses a volumetric syringe moved by a stepper motor that draws and delivers a known sample volume. Therefore, its absolute volumetric counting allows the direct determination of the number of cells per unit of sample volume without the need for reference material such as micro beads [8, 22]. Thus, direct volumetric CD4 T cell measurements were performed on the Auto40 using PE-conjugated anti-CD4 and PE-Dyomics649-conjugated anti-CD45 monoclonal antibodies (Apogee Flow Systems Ltd) for CD4 T cell count measurement. Anti-CD4 and anti-CD45 monoclonal antibodies are resistant to high temperatures because of their stabilization by a special dehydration process under controlled conditions (proprietary procedure, Bio-D) [11]. The Auto40 analytical procedure avoids the need for a wash step. Briefly, 50 μl of whole EDTA-blood was added into polypropylene test tubes containing pre-dispensed, stabilized monoclonal antibodies. After 25 minutes of incubation at room temperature in the dark, 450 μl of phosphate buffered saline were added. The no-lyse-no-wash stained samples were run on the Auto40 flow cytometer, and CD4 T cell count was obtained in absolute number and in percentage. Analysis on the Auto40 flow cytometer is automatically performed by the built-in very friendly software "Auto-Lymphocyte" (Apogee Flow Systems Ltd), with the possibility of controlling and assessing the quality of the data analysis, particularly when the CD4 T cell number is less than 40 CD4 T cells/μl. Aside from simplifying sample preparation, this system greatly facilitates manipulations by lab technicians. The training of technicians to use the Auto40 analyzer takes roughly one day.

Assessment of the precision of CD4 T cell counting in the mobile unit

The precision of CD4 T cell counting on the Auto40 instrument in the mobile unit was assessed using 2 different blood samples. Intrarun (instrument) precision was determined by preparing a large volume (10 ml) of CD4-stained blood and repeating sample acquisition on the instrument at least 10 times to determine the instrument's precision. Interrun precision, which includes the variation induced by pipetting errors made by the technician (tube-to-tube variability), was assessed by repeating the entire CD4 staining procedure at least 10 times: pipetting, sample preparation, staining, and sample acquisition. By comparing the intrarun and interrun precision, a fairly good impression of instrument's and technician's performance can be obtained. Interperson variation (between different technicians) was not assessed. Precision was expressed as the coefficient of variance (CV) obtained by dividing the standard deviation (SD) of all the measurements by the mean (CV% = SD × 100/mean).

Statistical analyses

The Method Validator software, version 1.1.9.0. (Philippe Marquis, France) and the SAS-PC software (version 8.2, SAS Institute, Cary, North Carolina, USA) were used for analyses. Firstly, correlations between the absolute CD4 T cell counts obtained by the reference FACSCalibur and the mobile Auto40 were established by the Passing-Bablok method which, in common with all nonparametric methods, is less sensitive to outliers [23]. Secondly, the agreement between the two methods was illustrated by difference plots as proposed by Bland and Altman [24, 25]. The average difference between the 2 methods, referred to as bias, was marked on the graph by a horizontal line, and the mean difference and the limits of agreement with a 95% confidence interval (CI) were also depicted.

To assess the clinical impact of using the Auto40 instead of the FACSCalibur in this setting, the sensitivity and the specificity of the Auto40 was calculated to identify patients who had with the FACSCalibur a CD4 T cell count below 200 cells/μl the threshold of immune-restoration under ART and the threshold for therapeutic initiation according to the 2006-revised WHO recommendations [26] or 350 cells/μl, the new threshold for ART initiation according to the 2010-revised WHO guidelines [1]. For clinical significance of the measurement differences on treatment decision, the Cohen's k coefficient was calculated on the whole study population [27].

Results

Accuracy of direct volumetric CD4 T cell measurements on the Auto40 using EDTA-containing whole blood

Study volunteers were included irrespective of their HIV sero-status (except children). The campaign in the suburb of Yaoundé included 98 adults volunteers (median age, 39.5 years; 28 males) for CD4 T cell count, and 93 children less than 5 years and more than 18 months (43 males). The campaign in Ambang Bikok included 69 adult volunteers (median age, 34 years; 39 males). Three blood samples from adult volunteers showed poor pre-analytical preparation when arriving at CIRCB, and were excluded from further analysis. Finally, parallel CD4 T cell measurements on both instruments, the Auto40 and FACSCalibur, were available for a total of 257 tested blood samples (Table 1).

Table 1 CD4 T cell counting in Yaoundé suburb and Ambang Bikok village obtained by the Auto40 flow cytometer installed in a therapeutic mobile unit, and by the FACSCalibur at the reference laboratory, expressed in 5th-95th percentile and mean.

In the suburb of Yaoundé, mean ± SD CD4 T cell count was 996 ± 874 cells/μl (range, 5-4295) by Auto40, and 989 ± 883 cells/μl (range, 5-4454) by FACSCalibur. In Ambang Bikok, mean ± SD CD4 T cell count was 1041 ± 317 cells/μl (range, 236-1774) by Auto40, and 1032 ± 294 cells/μl (281-1616) by FACSCalibur. The differences between Auto40 and FACSCAlibur were not statistically significant in suburb of Yaoundé as in Ambang Bikok (P > 0.5).

Results by Auto40 and FACSCalibur were highly correlated using a non-parametric Passing-Bablok regression analysis in Yaoundé (r2 = 0.98, slope = 1.01, intercept = 8.1) (Figure 1, top) as in Ambang Bikok (r2 = 0.92, slope = 1.07, intercept = -68) (Figure 1, bottom). The relation between Auto40 and FACSCalibur did not differ from linearity in suburb of Yaoundé as in Ambang Bikok (P > 0.1).

Figure 1
figure 1

Results from the campaigns in the suburb of Yaoundé (top) and in the village of Ambang Bikok (bottom), during which 188 and 69 CD4 T cell count measurements, respectively, were carried out in parallel by the Auto40 flow cytometer in mobile unit and by FACSCalibur in the Centre International de Référence Chantal Biya, Yaoundé. - Left. Passing-Bablok agreement test between the CD4 T cell count results expressed in cells/μl obtained with the Auto40 and FACSCalibur. The diagonal dotted line represents the idealline (no bias). The full line represents the regression line of the distribution; - Right. Bland-Altman analysis on the relative differences between the absolute CD4 T cell counts obtained with the Auto40 and FACSCalibur flow cytometers compared with the average absolute CD4 T cell count. The black full line represents the mean relative difference, and the other broken lines represent the superior and inferior limits of agreement. The arrow corresponds to the X abscise axis.

In the suburb of Yaoundé, the Auto40 results showed a bias of +6.8 cells/μl (95% CI: -12.4 - 26.0) as compared with FACSCalibur (Figure 1, top). In the village Ambang Bikok, the Auto40 results showed a bias of +9.7 cells/μl (95% CI: -19.9 - 39.4) as compared with FACSCalibur (Figure 1, bottom). Thus, Bland-Altman analysis on the relative differences between the absolute CD4 T cell counts obtained with Auto40 and FACSCalibur with the average absolute CD4 T cell counts results in suburb of Yaoundé and Ambang Bikok showed a close agreement between both methods. When pooling the 257 CD4 T cell count measurements, the Auto40 yielded a mean difference of +7.6 CD4 T cells/μl higher than by reference flow cytometry.

Analysis of CD4 T cell count measurement expressed in percentage showed, similarly to CD4 T cell count expressed in absolute numbers, a high correlation and a close agreement between both CD4 T cell counting methods in Yaoundé as in Ambang Bikok (data not shown).

The Figure 2 depicts the results from 93 CD4 T cell count measurements from children aged less than 5 years and more than 18 months, expressed in absolute number and in percentage. In this pediatric population, analysis of CD4 T cell count measurement expressed in percentage showed, similarly to CD4 T cell count expressed in absolute numbers, a high correlation and a close agreement between both CD4 T cell counting methods. Mean ± SD CD4 T cell count in percentage was 32.7 ± 12.1%CD4 (range, 10-70) by Auto40, and 33.0 ± 12.8%CD4 (range, 9-75) by FACSCalibur. The differences between Auto40 and FACSCAlibur were not statistically significant (P > 0.5). Results by Auto40 and FACSCalibur were highly correlated by regression analysis (r2 = 0.986, slope = 1.00, intercept = 0.0) (Figure 2, C). The relation between Auto40 and FACSCalibur did not differ from linearity (P > 0.5). The Auto40 results showed a bias of -0.30%CD4+ (95% CI: -0.91 - 0.31) as compared with FACSCalibur (Figure 2, C), thus demonstrating a close agreement between both methods to measure CD4 T cells in percentage.

Figure 2
figure 2

Results of 93 CD4 T cell count measurements from children aged less than 5 years and more than 18 months in the suburb of Yaoundé, expressed in absolute number (cells/μl) and in percentage carried out in parallel by the Auto40 flow cytometer and by FACSCalibur. - A and C. Passing-Bablok agreement test between the CD4 T cell count results obtained with the Auto40 and FACSCalibur flow cytometers, in absolute number (A) and in percentage (C). The diagonal dotted line represents the ideal line (no bias). The full line represents the regression line of the distribution; - B and D. Bland-Altman analysis on the relative differences between the CD4 T cell counts obtained with the Auto40 and FACSCalibur flow cytometers compared with the average CD4 T cell count, in absolute number (B) and in percentage (C). The full line represents the mean relative difference, and the dotted lines represent the superior and inferior limits of agreement. The arrow corresponds to the X abscise axis.

Sensitivity and specificity to identify clinically-relevant thresholds by the Auto40 installed in mobile unit

The sensitivity and specificity of primary CD4 T cell counting on the Auto40 in mobile unit to identify patients having less than (or more than) 200 CD4 T cells/μl was evaluated on the 257 available CD4 T cell count measurements. Taking into account a 10% bilateral range (i.e., counts between 190 and 210 CD4 T cells/μl were considered similar), the concordance between the Auto40 and FACSCalibur methods was high (k = 0.98; P < 0.01). The decision differed for 2 study individuals. Accordingly, the Auto40 had a sensitivity of 87% and a specificity of 99% to identify individuals with CD4 T cell counts below 200 cells/μl when compared with the FACSCalibur results.

The sensitivity and specificity of primary CD4 T cell counting on the Auto40 in mobile unit to identify patients having less than (or more than) 350 CD4 T cells/μl was also evaluated on the 257 available CD4 T cell count measurements. Considering a 10% bilateral range (i.e., counts between 332 and 367 CD4 T cells/μl were considered similar), the decision did not differ between study according to both methods (k = 0.96). The Auto40 had a sensitivity of 87% and a specificity of 98% to identify individuals with CD4 T cell counts below 350 cells/μl when compared with the FACSCalibur results.

Precision of direct volumetric CD4 T cell measurements on the Auto40 in mobile unit

The intra- and inter- run precisions were tested using 2 blood samples. The intrarun precision of Auto40 flow cytometer, expressed as the coefficient of variance, was 3.7% for the first sample (mean ± SD: 425 ± 16 CD4 T cells/μl) and 7.4% for the second sample (122 ± 9 CD4 T cells/μl), giving a mean intrarun precision of 5.5%. The interrun precision performed on the same 2 samples was 6.1% (439 ± 27 CD4 T cells/μl) and 9.7% (133 ± 13 CD4 T cells/μl), respectively, giving a mean interun precision of the Auto40 in mobile unit of 7.9%.

Discussion

In the present study, we demonstrated that CD4 T cell counting can be integrated into a therapeutic mobile unit and that the Auto40 flow cytometer, whether operated by a laboratory technician, performs acceptably compared with the FACSCalibur for absolute as well percentage count CD4 T cell testing. Thus, the essential WHO recommended ART staging and monitoring diagnostic tests can be accurately conducted at therapeutic mobile unit level using mini flow cytometer to accurately count CD4 T cells. The Auto40 flow cytometer installed in a mobile unit and operated far from its reference laboratory gave a perfect correlation with the reference method. The correlation was maintained over all the dynamic range of values (5-4295 CD4 T cells/μl). CD4 T cell counting in the therapeutic mobile unit permitted identification of the majority of individuals with CD4 T cells below 200 cells/μl, with a sensitivity of 87% and a specificity of 99%, and all individuals with CD4 T cells below 350 cells/μl, with a sensitivity of 87% and a specificity of 98%, demonstrating the capacity of Auto40 in mobile unit to measure in clinical practice CD4 T cells around the threshold of immune-restoration (200 cells/μl), and around the new threshold for therapeutic initiation (350 cells/μl) according to the 2010-revised WHO recommendations [1]. The procedure was fast, and needed only 30 minutes to be completed, with a capacity of 120 tests per day in routine. The technique was found to be easy to carry out and highly reproducible, with intra- and inter- run precisions less than 10%, considered as acceptable for clinical use [10]. Taken together, our observations address the proof-of-concept of point-of-care CD4 monitoring by therapeutic mobile unit, and support the decentralization of ART in rural settings with limited laboratory infrastructure. Indeed, the sturdiness of the Auto40 system associated with the use of heat-resistant reagents make possible the immunological monitoring of HIV-infected adults and children living in areas without laboratory facilities.

One of the main features of the Auto40 assay is the use of stabilized monoclonal antibodies. The assay reagents can be stored for prolonged period of time (up to 12 months) at high temperature without any loss of biological activity [11]. This feature is of potential application to the screening of HIV-infected individuals in resource-poor settings where conditions for the storage of reagents, particularly during shipment and delivery, cannot always be guaranteed. The thermo-stable monoclonal antibodies we used in the present study can be kept as long as one year at room temperature, i.e. up to +30°C, the stability at higher temperature being unknown. The possibility of long-term storage of reagents at room temperature should facilitate the planning of laboratory activities and reduce the costs related to loss of reagents. Overall, the use of thermo-stable reagents increases the accessibility to flow cytometry testing, since the increase in manufacturing costs related to the antibody stabilization procedure does not exceed 15% of the original cost [11].

The successful scale-up of ART will depend on the ability to deliver health services closer to patients and the community, in primary health care settings [2830], not only to reach more patients, but also to improve patient outcomes, including higher retention rates [31] and better virological suppression [3236]. The UNAIDS ART 2.0 model envisions decentralized ART that delivers treatment and care easily and inexpensively, close to where patients live [37]. Decentralization of diagnostics, which are typically conducted at district or central laboratories remote from primary health care settings, is critical to this approach. Patients often make several clinic visits to complete the testing process and results may take days or weeks to return. As transport costs and logistical challenges are significant causes of loss-to-follow-up [38, 39], tests conducted closer to the patient may improve retention and treatment success.

A wide assortment of more cost-efficient and technically less complex analyzers using proven flow cytometry-based technology have been developed in response to the overarching challenge of increasing access to CD4 T cell enumeration, and are especially suitable for resource-constrained settings [40]. Affordable CD4 T cell measurement, in absolute number and percentage, has gradually been possible by using simple, compact and robust low-cost new generation flow cytometers operating as single-platform volumetric instruments without the use of expensive micro beads [9]. Currently available mini flow cytometers present convenient efficiency when performed by well-trained laboratory technicians and when combined with good sample transport systems [9, 40]. In addition, several other point-of-care CD4 testing options have been conceived, and some are already on the market [41], all in a bid to improve access to CD4 T cell monitoring, especially for rural patients, and to reduce loss-to-follow-up of patients [40, 41].

The implementation of CD4 T cell counting in therapeutic mobile unit may help improve access to ART for patients in need, especially for rural populations and populations accessible mainly through mobile and outreach health services [15]. Previous experience on mobile health units for expansion of antiretroviral treatment to remote populations have been reported in Zambia [42]. The mobile service increased the number of ART clients in the study district probably because it reduced the long distances required to travel to health services in rural areas. Interestingly, clients found HIV-infected in mobile unit were generally starting ART at an early stage of HIV disease [42], likely because mobile ART services might have also encouraged people to seek voluntary counseling and testing, much earlier before presenting symptoms. Furthermore, involvement of the community such as lay counselors and support groups increased the number of patients retained at the original site compared to hospitals [42]. Thus, in the Zambian experience, transfers and lost-to-follow-up patients at the mobile sites during the first six months of ART were less frequent, with slightly higher retention rates (70-76%) [42] than generally observed in African hospital-based services (60%) [32]. Although HIV screening in mobile unit was demonstrated to be highly accepted in Cameroonese populations [15], the risk of possible stigma should be always be prevented. Taken together, mobile ART services in resource-limited settings can increase the number of ART patients by reducing the need to travel long distances to reach health facilities, as well as enable patients to start ART at an earlier stage in their disease when HIV diagnosis in made in voluntary counseling and testing center.

Monitoring of CD4 T cell count is cost-effective and may be cost-saving when compared to clinical monitoring alone for determining the timing of ART initiation, way before the development of life-threatening symptoms [4345]. The issue of whether the monitoring of patients on ART with a mobile unit is a valid economic option remains unresolved. Mobile health clinics are an appealing means of conducting outreach in that they provide an appropriate setting for sexually transmitted disease testing as well as HIV counseling, testing, and referral services. Mobile health vans are expensive, however, frequently costing two to three times more per patient compared with traditional clinics [46, 47]. In Uganda, the mobile clinic care for provision of ART has shown to be less cost-effective than facility-based care with higher incremental cost-effectiveness ratio, but would be competitive in targeted "hard-to-reach" population excluded from hospital-fixed health care [48].

Conclusions

The Auto40 flow cytometer installed in a therapeutic mobile unit and operated far from its reference laboratory gave a perfect correlation with the reference method, could be useful in carrying out immunological monitoring of HIV-infected patients living in areas without access to laboratory facilities, and could likely contribute to the decentralization and scaling-up of ART in Cameroon.

References

  1. World Health Organization: World Health Organization (WHO) recommendations Antiretroviral therapy for HIV infection in adults and adolescents. Antiretroviral therapy for a public health approach 2010 revision. [http://whqlibdoc.who.int/publications/2010/9789241599764_eng.pdf]

  2. World Health Organization: World Health Organization (WHO) recommendations. Antiretroviral therapy for HIV infection in infants and children: toward universal access. [http://whqlibdoc.who.int/publications/2010/9789241599801_eng.pdf]

  3. Bélec L, Bonn JP: Challenges in implementing HIV laboratory monitoring in resource-constrained settings: how to do more with less. Future Microbiol. 2011, 6: 1251-1260. 10.2217/fmb.11.121.

    Article  PubMed  Google Scholar 

  4. WHO laboratory recommendations by level. [http://www.who.int/hiv/amds/WHOLabRecommendationBylevelFinal.pdf]

  5. WHO consultation on technical and operational recommendations for scale-up of laboratory services and monitoring HIV antiretroviral therapy in resource-limited settings. [http://www.who.int/hiv/pub/meetingreports/scaleup/en/]

  6. Diagnostics and laboratory technology (DLT) website. [http://www.who.int/diagnostics_laboratory/en/]

  7. Mandy F, Nicholson J, McDougal J, CDC: Guidelines for performing single-platform absolute CD4R T-cell determinations with CD45 gating for persons infected with human immunodeficiency virus. Centers for Disease Control and Prevention. MMWR Recomm Rep. 2003, 52 (RR-2): 1-13.

    PubMed  Google Scholar 

  8. Dieye TN, Vereecken C, Diallo AA, Ondoa P, Diaw PA, Camara M, Karam F, Mboup S, Kestens L: Absolute CD4 T-cell counting in resource-poor settings: direct volumetric measurements versus bead-based clinical flow cytometry instruments. J Acquir Immune Defic Syndr. 2005, 39: 32-37. 10.1097/01.qai.0000160515.20581.ad.

    Article  PubMed  Google Scholar 

  9. O'Gorman MR, Zijenah LS: CD4 T cell measurements in the management of antiretroviral therapy--a review with an emphasis on pediatric HIV-infected patients. Cytometry B Clin Cytom. 2008, 74: S19-S26.

    Article  PubMed  Google Scholar 

  10. Dieye TN, Diaw PA, Daneau G, Wade D, Niang MS, Camara M, Diallo AA, Kane CT, Ndiaye HD, Mbengue B, Dieye A, Kestens L, Mboup S: Evaluation of a flow cytometry method for CD4 T cell enumeration based on volumetric primary CD4 gating using thrermoresistant reagents. J Immunol Methods. 2011, 372: 7-13. 10.1016/j.jim.2011.07.012.

    Article  CAS  PubMed  Google Scholar 

  11. Barbesti S, Soldini L, Carcelain G, Guignet A, Colizzi V, Mantelli B, Corvaglia A, Tran-Minh T, Dorigatti F, Autran B, Lazzarin A, Beretta A: A simplified flow cytometry method of CD4 and CD8 cell counting based on thermoresistant reagents: implications for large scale monitoring of HIV-infected patients in resource-limited settings. Cytometry B Clin Cytom. 2005, 68: 43-51.

    Article  PubMed  Google Scholar 

  12. UNAIDS: Report on the global AIDS epidemic 2010. [http://www.unaids.org/globalreport/Global_report.htm]

  13. Taiwo BO, Murphy RL: Clinical applications and availability of CD4 T cell count testing in sub-Saharan Africa. Cytometry B Clin Cytom. 2008, 74B (Suppl 1): S11-S18.

    Article  Google Scholar 

  14. Larsen CH: The fragile environments of inexpensive CD4 T-cell enumeration in the least developed countries: strategies for accessible support. Cytometry B Clin Cytom. 2008, 74B (Suppl 1): S107-S116.

    Article  Google Scholar 

  15. Mbopi-Kéou FX, Ongolo-Zogo P, Angwafo F, Ndumbe PM, Bélec L: High impact of mobile units for mass HIV testing in Africa. AIDS. 2007, 21: 1994-1996. 10.1097/QAD.0b013e3282f006c3.

    Article  PubMed  Google Scholar 

  16. O'Connor CA, Patsdaughter CA, Grindel CG, Taveira PF, Steinberg JL: A mobile HIV education and testing program: bringing services to hard-to-reach populations. AIDS Patient Care STDS. 1998, 12: 931-937. 10.1089/apc.1998.12.931.

    Article  PubMed  Google Scholar 

  17. Vidal L: Disclosure and counselling: elements for a review of the literature about African experiences [in French]. Psychopathol Africaine. 1994, XXVI: 155-188.

    Google Scholar 

  18. Kranzer K, Wood R, Bekker LG: Earlier HIV diagnosis--are mobile services the answer?. S Afr Med J. 2010, 100: 671-674.

    Article  PubMed  Google Scholar 

  19. Grabbe KL, Menzies N, Taegtmeyer M, Emukule G, Angala P, Mwega I, Musango G, Marum E: Increasing access to HIV counseling and testing through mobile services in Kenya: strategies, utilization, and cost-effectiveness. J Acquir Immune Defic Syndr. 2010, 54: 317-323. 10.1097/QAI.0b013e3181ced126.

    Article  PubMed  PubMed Central  Google Scholar 

  20. Hoffman RA, Hansen WP: Simple and rapid measurement of human T lymphocytes and their subclasses in peripheral blood. Proc Natl Acad Sci USA. 1980, 77: 4914-4917. 10.1073/pnas.77.8.4914.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  21. Brando B, Barnett D, Janossy G, Mandy F, Autran B, Rothe G, Scarpati B, D'Avanzo G, D'Hautcourt JL, Lenkei R, Schmitz G, Kunkl A, Chianese R, Papa S, Gratama JW: Cytofluorometric methods for assessing absolute numbers of cell subsets in blood. Cytometry. 2000, 42: 327-346. 10.1002/1097-0320(20001215)42:6<327::AID-CYTO1000>3.0.CO;2-F.

    Article  CAS  PubMed  Google Scholar 

  22. Mercolino TJ, Connelly MC, Meyer EJ, Knight MD, Parker JW, Stelzer GT, DeChirico G: Immunologic differentiation of absolute lymphocyte count with an integrated flow cytometric system: a new concept for absolute T cell subset determination. Cytometry. 1995, 22: 48-59. 10.1002/cyto.990220110.

    Article  CAS  PubMed  Google Scholar 

  23. Passing H, Bablok W: A new biometrical procedure for testing the equality of measurements from two different analytical methods. Application of linear regression procedures for method comparison studies in clinical chemistry, Part I. J Clin Chem Clin Biochem. 1983, 21: 709-720.

    CAS  PubMed  Google Scholar 

  24. Bland JM, Altman DG: Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986, 1: 307-310.

    Article  CAS  PubMed  Google Scholar 

  25. Bland JM, Altman DG: Measuring agreement in method comparison studies. Stat Methods Med Res. 1999, 8: 135-160. 10.1191/096228099673819272.

    Article  CAS  PubMed  Google Scholar 

  26. World Health Organization: Antiretroviral therapy for HIV treatment of adults and adolescents: recommendations for a public health approach 2006 revision. [http://www.who.int/hiv/pub/guidelines/adult/en/index.html]

  27. Cohen J: A coefficient of agreement for nominal scales. Educ Psychol Meas. 1960, 20: 37-46. 10.1177/001316446002000104.

    Article  Google Scholar 

  28. Fredlund VG, Nash J: How far should they walk? Increasing antiretroviral therapy access in a rural community in northern KwaZulu-Natal, South Africa. J Infect Dis. 2007, 196 (Suppl 3): S469-S473.

    Article  PubMed  Google Scholar 

  29. Loubiere S, Boyer S, Protopopescu C, Bonono CR, Abega SC, Spire B, Moatti JP: Decentralization of HIV care in Cameroon: increased access to antiretroviral treatment and associated persistent barriers. Health Policy. 2009, 92: 165-173. 10.1016/j.healthpol.2009.03.006.

    Article  PubMed  Google Scholar 

  30. Wools-Kaloustian KK, Sidle JE, Selke HM, Vedanthan R, Kemboi EK, Boit LJ, Jebet VT, Carroll AE, Tierney WM, Kimaiyo S: A model for extending antiretroviral care beyond the rural health centre. J Int AIDS Soc. 2009, 12: 22-10.1186/1758-2652-12-22.

    Article  PubMed  PubMed Central  Google Scholar 

  31. Rosen S, Fox MP, Gill CJ: Patient retention in antiretroviral therapy programs in sub-Saharan Africa: a systematic review. PLoSMed. 2007, 4: e298-

    Google Scholar 

  32. Bedelu M, Ford N, Hilderbrand K, Reuter H: Implementing antiretroviral therapy in rural communities: the Lusikisiki model of decentralized HIV/AIDS care. J Infect Dis. 2007, 196 (Suppl 3): S464-S468.

    Article  PubMed  Google Scholar 

  33. Boyer S, Eboko F, Camara M, Abé C, Nguini ME, Koulla-Shiro S, Moatti JP: Scaling up access to antiretroviral treatment for HIV infection: the impact of decentralization of healthcare delivery in Cameroon. AIDS. 2010, 24: S5-S15.

    Article  PubMed  Google Scholar 

  34. Chan AK, Mateyu G, Jahn A, Schouten E, Arora P, Mlotha W, Kambanji M, van Lettow M: Outcome assessment of decentralization of antiretroviral therapy provision in a rural district of Malawi using an integrated primary care model. Trop Med Int Health. 2010, 15 (Suppl 1): 90-97.

    Article  PubMed  Google Scholar 

  35. Fatti G, Grimwood A, Bock P: Better antiretroviral therapy outcomes at primary healthcare facilities: an evaluation of three tiers of ART services in four South African provinces. PLoS One. 2010, 5: e12888-10.1371/journal.pone.0012888.

    Article  PubMed  PubMed Central  Google Scholar 

  36. Geng EH, Nash D, Kambugu A, Zhang Y, Braitstein P, Christopoulos KA, Muyindike W, Bwana MB, Yiannoutsos CT, Petersen ML, Martin JN: Retention in care among HIV infected patients in resource-limited settings: emerging insights and new directions. Curr HIV/AIDS Rep. 2010, 7: 234-244. 10.1007/s11904-010-0061-5.

    Article  PubMed  PubMed Central  Google Scholar 

  37. UNAIDS outlook report 2010. [http://data.unaids.org/pub/outlook/2010/20100713_outlook_report_web_en.pdf]

  38. Hardon AP, Akurut D, Comoro C, Ekezie C, Irunde HF, Gerrits T, Kglatwane J, Kinsman J, Kwasa R, Maridadi J, Moroka TM, Moyo S, Nakiyemba A, Nsimba S, Ogenyi R, Oyabba T, Temu F, Laing R: Hunger, waiting time and transport costs: time to confront challenges to ART adherence in Africa. AIDS Care. 2007, 19: 658-665. 10.1080/09540120701244943.

    Article  CAS  PubMed  Google Scholar 

  39. Miller CM, Ketlhapile M, Rybasack-Smith H, Rosen S: Why are antiretroviral treatment patients lost to follow-up? A qualitative study from South Africa. Trop Med Int Health. 2010, 15 (Suppl 1): 48-54.

    Article  PubMed  PubMed Central  Google Scholar 

  40. Murtagh MM: UNITAIDS Technical Report. HIV/AIDS diagnostic landscape. 2011, [http://www.unitaid.eu/images/marketdynamics/publications/unitaid_md_technical_report_diagnostics_landscape_web.pdf]

    Google Scholar 

  41. Jani IV, Sitoe NE, Chongo PL, Alfai ER, Quevedo JI, Tobaiwa O, Lehe JD, Peter TF: Accurate CD4 T-cell enumeration and antiretroviral drug toxicity monitoring in primary healthcare clinics using point-of-care testing. AIDS. 2011, 25: 807-812. 10.1097/QAD.0b013e328344f424.

    Article  PubMed  Google Scholar 

  42. Dube C, Nozaki I, Hayakawa T, Kakimoto K, Yamada N, Simpungwe JB: Expansion of antiretroviral treatment to rural health centre level by a mobile service in Mumbwa district, Zambia. Bull World Health Organ. 2010, 88: 788-791. 10.2471/BLT.09.063982.

    Article  PubMed  PubMed Central  Google Scholar 

  43. Goldie SJ, Yazdanpanah Y, Losina E, Weinstein MC, Anglaret X, Walensky RP, Hsu HE, Kimmel A, Holmes C, Kaplan JE, Freedberg KA: Cost-effectiveness of HIV treatment in resource-poor settings: the case of Cote d'Ivoire. N Engl J Med. 2006, 355: 1141-1153. 10.1056/NEJMsa060247.

    Article  CAS  PubMed  Google Scholar 

  44. Loubiere S, Meiners C, Sloan C, Freedberg KA, Yazdanpanah Y: Economic evaluation of ART in resource-limited countries. Curr Opin HIV AIDS. 2010, 5: 225-231. 10.1097/COH.0b013e3283384a9d.

    Article  PubMed  PubMed Central  Google Scholar 

  45. Walensky P, Ciaranello CL, Park J-E, Freedberg KA: Cost-effectiveness of laboratory monitoring in sub-Saharan Africa: A review of the current literature. Clin Infect Dis. 2010, 51: 85-82. 10.1086/653119.

    Article  PubMed  PubMed Central  Google Scholar 

  46. DiFranceisco W, Holtgrave DR, Hoxie N, Reiser WJ, Resenhoeft R, Pinkerton SD, Vergeront J: HIV seropositivity rates in outreach-based counseling and testing services: program evaluation. J Acquir Immune Defic Syndr Hum Retrovirol. 1998, 19: 282-288. 10.1097/00042560-199811010-00011.

    Article  CAS  PubMed  Google Scholar 

  47. Ellen JM, Bonu S, Arruda JS, Ward MA, Vogel R: Comparison of clients of a mobile health van and a traditional STD clinic. J Acquir Immune Defic Syndr. 2003, 32: 388-393. 10.1097/00126334-200304010-00007.

    Article  PubMed  Google Scholar 

  48. Babigumira JB, Sethi AK, Smyth KA, Singer ME: Cost effectiveness of facility-based care, home-based care and mobile clinics for provision of antiretroviral therapy in Uganda. Pharmacoeconomics. 2009, 27: 963-973. 10.2165/11318230-000000000-00000.

    Article  PubMed  PubMed Central  Google Scholar 

Download references

Acknowledgements

The authors are grateful to the Laboratoire National de Santé Hygiène Mobile, Yaoundé, Cameroon, for making this evaluation possible and to the Centre International de Référence Chantal Biya, Yaoundé, Cameroon, for performing the analyses on the FACSCalibur. We thank Rose Guiadem Kamdem, Chimi Laure and Mboubi-Massepo Yves Martial for their technical assistance. We also thank Jean-Louis Languillat, Inodex, France, and Oliver Kenyon, Apogee Flow Systems Ltd, United Kingdom, for their fruitful technical input, comments and discussions.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to François-Xavier Mbopi-Kéou.

Additional information

Competing interests

The authors declare that they have no competing interests.

Authors' contributions

FXMK, BS, JN, FFA, LB have conceived and designed the research; FXMK, BS have performed the experiments; LB performed statistical analyses; VC, LM help analyzed the results; FXMK, LB drafted the manuscript. All authors read and approved the final manuscript.

Authors’ original submitted files for images

Below are the links to the authors’ original submitted files for images.

Authors’ original file for figure 1

Authors’ original file for figure 2

Rights and permissions

This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Reprints and permissions

About this article

Cite this article

Mbopi-Kéou, FX., Sagnia, B., Ngogang, J. et al. Validation of a single-platform, volumetric, flow cytometry for CD4 T cell count monitoring in therapeutic mobile unit. J Transl Med 10, 22 (2012). https://doi.org/10.1186/1479-5876-10-22

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/1479-5876-10-22

Keywords